“J. Mater. Chem. B” reports on the detection of tetrahydrocannabinol (Δ9-THC) using aerosol-jet-printed organic electrochemical transistors.

Product Introduction

Equipment Parameters

I. Core Research Summary

This work leverages Optomec’s aerosol jet printing system to fabricate complete OECT devices on flexible Kapton films, eliminating the need for photolithography or evaporation deposition. Silver source and drain electrodes, a PEDOT:PSS sensing channel, and a UV‑curable insulating layer are sequentially patterned and integrated. An external platinum gate electrode is employed to construct a THC electrochemical sensing platform, enabling performance validation in both pure water and artificial saliva systems. The key findings are as follows:
  • Complete integrated OECT fabrication system for all aerosols

Ultrasonic atomization of silver and PEDOT:PSS inks, pneumatic atomization of UV‑PDMS insulating ink, and single‑device layer‑by‑layer printing of multilayer structures; channel thickness of 560 nm, with a silver film resistivity as low as 5×10⁻⁶ Ω·cm, and device transconductance reaching 1.5 mS, exhibiting intrinsic signal amplification without the need for external amplification circuitry.

  • Ultra-wide linear detection range, ultra-low limit of detection

In a pure water matrix, the detection limit is 0.1 nM, with a measurement range spanning 0.1 nM to 5 mM; in an artificial saliva environment, the detection limit is 1 nM, fully encompassing the physiological THC concentrations in human saliva (3.18–1039.88 nM), exhibiting two‑segment linear response across low and high concentration ranges, with a measurement error of less than 3%.

  • Ultra-fast detection of trace samples, with extremely simple operation.

Only 2–10 μL of saliva is required, with stable signal acquisition lasting approximately 5 minutes and no complex sample preparation needed. Repeated measurements demonstrate high signal consistency, and a single device exhibits a cycle-to-cycle measurement deviation of less than 2.65% across seven cycles.

  • Stable recognition of complex salivary matrices

Even in the presence of multiple ions and organic interferents, artificial saliva can accurately distinguish THC concentrations, with only brief baseline fluctuations; once equilibrium is reached, the current response remains stable, demonstrating suitability for on-site sample analysis.

  • Low-cost, single-use sensor chips are scalable.

Digital maskless printing, combined with a flexible, lightweight, and portable substrate, enables rapid iteration of device architectures. Subsequent functionalization with THC antibodies can further enhance specificity, making it well-suited for the mass production of security‑screening consumables.

II. Complete Preparation Process and Core Advantages of Aerosol Spraying

(1) Complete OECT Aerosol Printing Process

  • 25.4 μm Kapton flexible film substrate cleaning and pre-treatment;
  • Ultrasonic atomization of silver nano‑ink, aerosol printing source / leakage electrode, followed by overnight annealing at 140°C to reduce resistivity;
  • Ultrasonic atomization–modified PEDOT:PSS composite ink, with a double-layer printed sensing channel, cured and formed at 130°C.
  • Pneumatically atomized and diluted UV‑PDMS insulating ink is used to precisely print the electrode passivation layer, followed by 405 nm UV‑induced simultaneous crosslinking to prevent electrolyte‑induced short circuits.
  • An electrochemical detection platform was assembled using an external platinum wire grid, and a saliva sample was dropwise added to acquire the THC signal.

(II) Five Major Process Advantages (Compared with Inkjet, Screen Printing, and Photolithography for OECT Fabrication)

  • Three-layer heterogeneous materials integrated into a single device via one-step molding.

Ultrasonic processing is compatible with conductive polymers and metal inks, while pneumatic processing works well with high-viscosity insulating adhesives. This enables the simultaneous fabrication of electrodes, sensing channels, and insulating protective layers, whereas photolithography requires multiple mask‑alignment steps, evaporation, and etching processes, significantly reducing the overall cycle time.

  • Micron-level precise partitioned deposition, with precisely defined openings in the insulating layer.

The printing area is digitally constrained, with PDMS coating only the conductive leads while leaving the sensing channels fully exposed. The electrolyte contacts only the functional regions, thereby eliminating leakage currents and inter-channel crosstalk and enhancing the signal-to-noise ratio.

  • Low-temperature, mild processing, compatible with flexible polymer substrates.

The maximum curing temperature is only 140°C, preventing warping and embrittlement of the PI film. The ultra-thin film can be laminated onto portable test‑strip carriers, making it ideal for disposable, adhesive‑type test cards.

  • Maskless Rapid Iterative Sensing Configuration

With CAD, the channel aspect ratio and electrode dimensions can be adjusted with a single click, enabling rapid optimization of OECT transconductance for different toxicant molecules; new screening chip prototypes can be completed in just a few hours.

  • Consumables are deposited on demand, resulting in low mass-production costs.

Only the sensing area is coated with noble metals and conductive polymers, resulting in a 70% material loss across the entire film; the flexible substrate is inexpensive and readily available, making it well suited for high‑volume, single‑use detection consumables in transportation and security applications.

III. Five Key Technological Innovations

Innovation 1: Innovative Preparation of an All-Aerosol Integrated THC Detection OECT

Breaking with the conventional OECT fabrication approach that relies on photolithography or inkjet printing, this work employs aerosol jet writing for all components—silver interconnects, PEDOT channel, and insulating passivation layer—enabling a fully flexible transistor‑based sensing platform suitable for electrochemical detection of biofluids.

Innovation 2: Leveraging a novel label-free sensing mechanism based on electrochemical oxidation of THC.

Without the need for enzymatic or antibody pre‑functionalization, the phenolic groups of THC generate a Faradaic current upon oxidation at the Pt gate, modulating the OECT gate voltage and thereby controlling the channel current. This approach simplifies chip‑modification procedures and reduces consumable costs.

Innovative 3 OECT endogenous signal amplification system, eliminating the need for an external amplification module.

Organic electrochemical transistors exhibit intrinsic transconductance, enabling the amplification of weak electrochemical signals. At comparable THC concentrations, their signal intensity far exceeds that of conventional three-electrode printed sensors, and even trace‑level saliva samples at low concentrations can be stably detected.

Innovation 4: A Hierarchical Linear Sensing Model Adapted to Complex Saliva Matrices

It distinguishes between low-concentration (1–425 nM) and high-concentration (425 nM–5 mM) dual-response ranges, aligns with the actual salivary concentration profile in humans, and maintains stable quantitative performance even in high-salt, multi-ion environments.

Innovation 5: A Universal Aerosol-Printing Platform for Standardizing Toxicity Sensing

By simply replacing the gate‑modifying biorecognition molecule, the platform can be extended to detect a wide range of bodily fluids, including alcohol, drugs, and metabolites, enabling a single additive manufacturing line to support the development of multiple types of point‑of‑care (POCT) chips.

IV. Existing Process Challenges and Corresponding Solutions for Overcoming Them

Challenge 1: The layered fabrication of multi-material OECTs is cumbersome, and conventional processes require multiple pieces of equipment.

✅ Breaks through the aerosol dual‑atomization module switching, enabling simultaneous printing of metals, conductive polymers, and insulating adhesives, and allowing a single machine to fabricate a complete transistor, thereby simplifying production line layout.

Challenge 2: Conventional printed electrodes produce weak signals, and low-concentration THC is easily masked by noise.

✅ Breaking through the self-amplifying architecture of OECT transistors, and integrating an aerosol‑derived, high‑surface‑area PEDOT porous channel, weak oxidation signals can be significantly amplified, with a detection limit as low as 0.1 nM.

Challenge 3: Saliva contains a large number of ionic impurities, which can easily lead to baseline drift and inaccurate quantification.

✅ Optimized aerosol printing of a thick, uniform PEDOT channel, enhancing ion‑buffering capacity; after sample equilibration, the baseline remains stable, and the concentration–current response exhibits excellent linearity.

Challenge 4: The lithography-based OECT prototyping process can take several days, making rapid iterative screening of chips impractical.

✅ Maskless digital aerosol direct writing enables on‑day fabrication of new devices and rapid validation of drug‑testing targets.

Challenge 5: The unmodified bare-gate exhibits insufficient specificity and is easily interfered with by similar phenolic compounds.

✅This process incorporates design flexibility for gate‑modifying modifications; subsequently, by employing aerosol‑based spot‑printing of THC antibodies, molecular recognition specificity can be significantly enhanced.

V. Future Diversified Industrial Application Scenarios

1. On-site rapid screening for driving under the influence of alcohol or drugs

Traffic police portable saliva test cards require no laboratory equipment, deliver results in 5 minutes, and accurately identify individuals who have consumed THC, serving as a substitute for laboratory‑based mass spectrometry testing and significantly enhancing on‑site enforcement efficiency.

2. Mass screening at airports, train stations, and border checkpoints

Single-use aerosol‑based test strips enable rapid initial screening of large passenger volumes, with suspicious samples subsequently sent to the laboratory for confirmatory testing, thereby reducing security‑check time and testing costs.

3. Long-term medication monitoring at drug rehabilitation facilities

Regular saliva sampling of individuals undergoing drug rehabilitation enables non-invasive monitoring of residual THC levels in the body, assisting clinicians in assessing withdrawal and recovery status and facilitating simple at-home monitoring.

4. Rapid Preliminary Screening in Forensic Toxicology

Rapid qualitative analysis of bodily fluids at crime scenes involving homicide or drug-related offenses reduces turnaround time for sample submission and shortens the investigation cycle.

5. A universal detection chip platform for multi-component body fluids

By replacing the gate‑surface bio‑modification material, the platform can be extended to detect alcohol, stimulants, tumor markers, and other analytes, making it compatible with portable biochemical screening devices used in primary care clinics and health‑checkup centers.
Image and Text Guide

Figure 1. (a) Microscopic image of an aerosol‑jet‑printed OECT, with a scale bar of 500 µm. The channel region is shown in dark blue (PEDOT:PSS), covering the silver electrode tip. (b) Potential drop across the two electrical double layers (EDLs) at the platinum gate electrode surface and the PEDOT:PSS channel surface; capacitances CGE and CCE are used as electrical models to characterize the EDLs. (c) Schematic diagram of a D9‑THC sensor based on the OECT. (d) Chemical structure of Δ9‑tetrahydrocannabinol (D9‑THC) and the electrooxidation process of the phenolic group in its molecule (references 66, 67). (e) Cyclic voltammograms obtained using a water‑diluted D9‑THC solution. (f) Plot of the anodic peak current (Ip,a) as a function of D9‑THC concentration; Ip,a values were extracted from data acquired at a potential of 0.26 V for each D9‑THC concentration shown in (e).

Figure 2 shows the drain current (ID) response curves of aerosol‑jet‑printed OECTs under a fixed bias condition of VD = 0.1 V and VG = 1 V, following the addition of D9‑THC at different concentrations. These measurements were performed on three distinct OECT devices: (a) and (b) correspond to different devices, while (c) through (f) were carried out on the same OECT after each measurement underwent a cleaning step to remove residual D9‑THC, thereby confirming the device’s reproducibility. (a) presents the ID response curves for solutions containing 0.1 nM, 10 nM, 500 nM, and 5 mM D9‑THC in deionized water, with the corresponding real-time drain currents indicated at each concentration. (b) displays the ID responses upon adding 1 nM, 5 nM, 10 nM, 100 nM, 1 mM, and 5 mM D9‑THC solutions. (c) illustrates the ID responses when 0.1 nM, 0.5 nM, 1 nM, and 2.5 nM D9‑THC solutions are introduced. (d) shows the ID responses for 2.5 nM, 5 nM, 10 nM, and 50 nM D9‑THC solutions. (e) presents the ID responses for 50 nM, 100 nM, 500 nM, and 1 mM D9‑THC solutions. (f) depicts the ID responses when 1 mM, 2.5 mM, and 5 mM D9‑THC solutions are applied.

Figure 3. Normalized drain current responses of the OECT device upon addition of D9‑THC solutions at different concentrations. (a) Average normalized drain current response curves for two OECT devices measured using D9‑THC solutions at 0.1 nM, 10 nM, 500 nM, and 5 mM: the blue series corresponds to seven replicate measurements, while the black series reflects four replicates. Both curves exhibit similar slope characteristics: the blue curve has a slope of 0.0153 per decade in the 0.1–500 nM range and 0.1300 per decade in the 0.5–5 mM range; the black curve shows a slope of 0.0134 per decade from 0.1 to 500 nM and 0.1276 per decade from 0.5 to 5 mM. (b) Average normalized drain current response curve of the OECT device measured with D9‑THC solutions at various concentrations (1 nM, 5 nM, 10 nM, 100 nM, 1 mM, and 5 mM): both curves display comparable operating‑range features—slope of 0.02 per decade from 1 to 573 nM and 0.122 per decade from 573 nM to 5 mM; all data presented were obtained from four replicate measurements.

Figure 4. (a) ID response curves of aerosol‑jet‑printed OECTs to synthetic saliva buffer upon addition of the buffer. A pronounced peak in the gate current is also observed when the synthetic saliva buffer is added, owing to the high concentration of cations in the solution. (b) ID response curves of aerosol‑jet‑printed OECTs to D9‑THC at various concentrations diluted in synthetic saliva buffer. (c) Normalized ID values of the aerosol‑jet‑printed OECTs after adding D9‑THC at different concentrations. The operating range is divided into two regions: a slope of 0.003 per decade from 1 to 425 nM, and a slope of 0.02 per decade from 425 nM to 5 mM.

Table 1 Comparison of Detection Time, Required Analyte Volume, Limit of Detection, and Sensitivity for Different Types of D9-THC Electrochemical Sensors